Model: DR-JT-S-200-10

Base temperature guaranteed less than 10mK
Cooling power at 100mK guaranteed greater than 200µW
Typical dimensions
- Outer diameter of the insert: 180 mm
- Total length of the insert with IVC: up to 2500 mm
- Approximate sample space at low temperature (depends on Dewar and magnet): from 140 mm in diameter up to 400 mm in length
Special Features:
- Joule Thomson exchanger (replacing a separately pumped bath condenser)
- Easily removable IVC (three clamps and an indium seal)
- Exchangeable connector plates on experimental lines
- Radiation shields and IVC can be designed to fit into a superconducting magnet
Optional Features:
- 1K pot for heatsinking of wiring or can be used for experiments at 1K
- Special access and/or line of sight ports
- Custom sample holder (Swedish rotators for example)
- Thermometry from 8mK to 4.2K
- Remote controlled Gas Handling System (GHS)

Temperature measurement is made with a TRMC2 controller and RuO2 thermometer calibrated by Nuclear Orientation. The temperature sensor is located in vacuum on the mixing chamber plate.
Download pdf drawing
Reference customers
|