HomeAboutContactCryomagnetics.com  

LHe-COOLED, 400µW @ 100mK, 10mK Base

Model: DR-JT-S-400-10

Base temperature guaranteed less than 10mK
Cooling power at 100mK guaranteed greater than 400µW

Typical dimensions:

  • Outer diameter of the insert: 250 mm
  • Total length of the insert with IVC: up to 2500
  • Sample space at low temperature (dependant on dewar and magnet): 200 mm in diameter and up to 400 mm in length

Special Features:

  • Joule Thomson exchanger (replacing a separate pumped bath condenser)
  • Easily removable IVC (Third clamps and indium seal)
  • Exchangeable connector plates on experimental lines
  • Radiation shields and IVC could be designed to fit into magnet

Optional Features:

  • 1K pot for heatsinking of wiring or can be used for experiments at 1K
  • Special access and/or line of sight ports
  • Custom sample holder (Swedish rotators for example)
  • Thermometry from 8mK to 4.2K
  • Remote controlled Gas Handling System (GHS) on request

Temperature measurements made with TRMC2 controller and RuO2 thermometer calibrated by Nuclear Orientation. Thermometer is located in vacuum on the mixing chamber plate.

Download pdf drawing

Reference customers